Indications for S therapy
•CPAP intolerance
•Continued obstructive events at
higher pressures
•Hypoventilation with SpO
2
< 90%
•COPD
Are events central?
For obstructive apneas:
•Increase EPAP by ≥1 cm H
2
O every ≥ 5 min
•Increase IPAP to maintain 4 cm H
2
O difference between IPAP/EPAP
Observe patient and document final settings,
including IPAP/EPAP pressures and TiControl
settings if altered from default
Has patient been on CPAP therapy?
Initial settings:
•IPAP = CPAP settings
•EPAP = 4 cm H
2
O
below IPAP setting
Initial settings:
•IPAP = 8 cm H
2
O
•EPAP = 4 cm H
2
O
For obstructive hypopneas and snoring:
•Increase IPAP ≥ 1 cm H
2
O every ≥ 5 min until resolved
For SpO
2
< 90% with all
respiratory events eliminated:
•Increase IPAP by ≥ 1 cm H
2
0 every
≥ 15 min until ≥ 90% SpO
2
is reached
•Follow sleep lab protocols for adding O
2
Decrease pressure
to previous setting,
observe for 20 min
If centrals persist,
consider adaptive
servo-ventilation
YES
YES
NO
NO
VPAP S Therapy Titration Protocol
(bilevel spontaneous)
Lab Support 24/ 7 (888) 288-6738 | Sleep Lab Titration Guide 35