Notes:
The numeric value (1, 2 or 3 cm H
2
O) for each EPR setting represents the maximum pressure
drop during CPAP therapy expiration. Therapy pressure will never drop below 4 cm H
2
O. So, for
example, if therapy pressure during Ramp Time is 5 cm H
2
O and EPR is set at level 3, then the
pressure will only reduce to 4 cm H
2
O.
For this reason, EPR is a unique comfort feature that ensures therapy effectiveness at all times.
EPR allows patient comfort without compromise because the selected setting offers a defined
pressure drop value that never exceeds the set value.
EPR
and Easy-Breathe
EPR is designed to maintain optimal treatment for the patient during inhalation and
reduce the delivered mask pressure during exhalation in the CPAP or AutoSet mode.
The desired result of EPR is to decrease the pressure the patient must breathe out
against, making the overall therapy more comfortable.
Comfort levels
EPR provides three comfort settings. Each comfort
setting correlates to an exact drop in pressure relief:
•EPR Level 1: Mild reduction (1 cm H
2
O)
•EPR Level 2: Medium reduction (2 cm H
2
O)
•EPR Level 3: Maximum reduction (3 cm H
2
O)
Device
Pressure
Patient
Flow
EPR Level 1
Mild pressure
EPR Level 2
Medium pressure
EPR Level 3
Maximum pressure
20 S9 VPAP Tx Lab System | Sleep Lab Titration Guide
ResMed.com
CPAP and AutoSet Technology