Time
Pressure
Pressure
Target alveolar ventilation
Actual alveolar ventilation
Time lapse
Pressure increases
to meet target alveolar
ventilation
Pressure decreases
to meet target alveolar
ventilation
Automatically changing pressure support
and iBR to maintain alveolar ventilation
Which patients is iVAPS suitable for?
iVAPS is suitable for adults with respiratory insufficiency. It
is ideal for patients whose condition is likely to change and
is characterized by hypoventilation (day/night hypercapnia).
Patient conditions may include:
•Neuromuscular disease and restrictive conditions iVAPS can
maintain stable ventilation when respiratory effort fluctuates
•Obesity hypoventilation When compared to standard
Pressure Support therapy, iVAPS can compensate for changes
in respiratory mechanics, such as during nocturnal changes in
the patient’s body position
•Chronic obstructive pulmonary disease iVAPS may reduce the
risk of hyperinflation associated with increased respiratory rate,
as compared to therapy targeting tidal volume
What does iVAPS target?
iVAPS targets alveolar ventilation to deliver required
ventilation at the alveoli, where gas exchange occurs.
Unlike other volume-assurance modes, iVAPS maintains
the alveolar target even when respiratory rate changes.
Why is alveolar ventilation important?
Setting alveolar ventilation targets the patient’s true
ventilation requirements and represents a more
accurate approach.
It’s able to deliver the required ventilation at the alveoli,
where gas exchange occurs, by taking into account and
compensating for the portion of air that travels through
the conducting airways.
What are the goals of therapy with iVAPS?
•Optimizing therapy by delivering a set alveolar ventilation with the
right pressure at the right time
•Enhancing patient-ventilator synchrony with an intelligent Backup
Rate (iBR) to enhance patient comfort
•Minimizing sleep disruption with its rapid, yet gentle response that
is quick enough to maintain stable alveolar ventilation, yet smooth
enough to maintain sleep quality during nocturnal therapy
•Increased adherence to therapy with iVAPS
1
iVAPS Technology Q&A
iBR maximizes the patient’s opportunity to breathe
spontaneously before bringing the patient back to
target if backup breaths are required
Patient
effort
No patient
effort
Patient
effort
iBR stays out of
the way when
not required
iVAPS intelligent backup rate (iBR)
Patient spontaneous rate
Target patient rate
Time
Breaths per minute
15
10
1 Jaye J, Kelly J et al. Eur Respir J 2012; 40: Suppl. 56, 51
26 S9 VPAP Tx Lab System | Sleep Lab Titration Guide
ResMed.com
iVAPS Technology